MPG500-CF275
Inficon MPG500 (Gemini) Series Combination Gauge
Pirani and Cold Cathode (Inverted Magnetron) Vacuum Gauge
Combination High Vacuum Full Range Gauge
• 2.75" OD Conflat Flange Port (Rotatable)
• Inficon PN: 3MB0-008-000P
• Voltage vs. Pressure: 0.6 V/decade (logarithmic)
High Vacuum Gauge, Measures From:
• 1x10-9 mbar to 1000
• 7.6x10-10 Torr to 760
MPG 500 Non-Surplus, Standard Inventory
Additional Technical Information On The Bottom Of This Page
2-3/4" OD CF Conflat Flange Equivalents
275CF, CF35, CF38, CF40, CF275
DN35CF, DN40CF, NW35CF
Drop In Replacement for Inficon:
• 351-012 MPG400, DN40CF (MPG400-CF275)
Drop In Replacement for Leybold Vacuum (Oerlikon):
• 230072 PTR90, DN40CF PenningVac
Drop In Replacement for Agilent (Varian Vacuum):
• FRG700CF35 FRG-700, 2.75" Conflat
P1012329 P103350 351-012 351012 351 012 Penning PTR 90
|
|
MPG500 Series
Specifications
|
|
|
Measurement range (air, N2)
|
1
x 10-9 … 1000 mbar
|
|
7.6
x 10-10 … 760 Torr
|
|
Accuracy (N2)
|
|
|
1
x 10-8 … 1 x 10-2 mbar
|
±30%
of reading
|
|
1
x 10-2 … 100 mbar
|
±30%
of reading
|
|
100
… 1000 mbar
|
±50%
of reading
|
|
Repeatability (N2)
|
±5%
of reading
|
|
1
x 10-8 … 100 mbar
|
|
Mounting
Orientation
|
ANY
|
|
Admissible pressure bar (absolute)
|
10
bar (limited to inert gases <55°C)
|
|
Admissible temperature
|
|
|
Operation (ambient)
|
+5
… +55°C
|
|
Storage
|
-40
… +70°C
|
|
Bake-out at Flange*
|
150°C
|
|
Filament temperature
(Pirani)
|
120°C
|
|
Supply Voltage
|
|
|
At Gauge
|
+14.5
… +30 VDC
|
|
Ripple
|
<1 V (pp)
|
|
Power
Consumption
|
<3 W
|
|
Fuse
to be connected
|
<1 AT
|
|
Output Signal
(Measurement Signal)
|
|
|
Voltage Range (analog
output)
|
0
… +10.5 V
|
|
Measurement Range
|
+1.398
… +8.598 V
|
|
Voltage vs. Pressure
|
logarithmic, 0.6 volts per decade
|
|
Response Time (pressure
dependent)
|
|
|
> 10-6 mbar
|
<100
ms
|
|
1 x 10-6 … 1 x 10-8 mbar
|
1
s
|
|
Identification gauge
|
85 kΩ
|
|
Status Signal (digital
output) FCC Connector
|
|
Current rating
|
100
mA
|
|
High voltage is ON
|
+14.5
… +30V (dc) depending on supply voltage
|
|
High Voltage is OFF
|
0
V (dc)
|
|
Electrical
Connection
|
FCC 68 appliance connector,
|
|
|
8 poles, female
|
|
Sensor
Cable
|
8 poles plus shielding
|
|
High
voltage (in the measuring chamber)
|
|
Ignition voltage
|
<4.5 kV
|
|
Operating Voltage
|
<3.3 kV
|
|
Current
in measuring chamber
|
|
High Current
|
<380 μA
|
|
Low Current
|
<130 μA
|
|
Materials exposed to vacuum
|
stainless
steel, Al2O3, Glass, Ti
|
|
|
Mo,
Ni alloy, W
|
|
*Without electronics
|
|
**The minimum voltage of the supply unit must be increased proportionally
to the length of the sensor cable.
|
Contact LDS Vacuum with any questions:
Phone (407)862-4643 via Fax (407)862-8723 or
info@ldsvacuum.com
This item is no longer available to customers with a FINAL DESTINATION OUTSIDE THE US, Mexico or Canada.
Contact Manufacturer To Purchase Directly.
LDS Vacuum Products is a 100% woman owned company and certified Woman Owned Small Business (WOSB) The logos and/or trademarks appearing on this page are not owned by or registered to LDS but, rather, designate products sold by LDS, and are owned and/or registered to the manufacturers or suppliers of the products. |
|